Features of the instrument
- Advanced sequential injection platform, greatly reduces the number of components, with a simple and easy-to-diagnose structure.
- High-precision and stable injection pump, non-contact liquid quantitative design, high dilution range with less water consumption and short dilution time.
- The ten-way valve flow path adopts laser micro-engraving technology to precisely control the size and positioning of the grooves. The valve core is made of sapphire material with ultra-high corrosion and wear resistance, long service life, and good stability.
- Integrated digestion/colorimetry unit with high absorbance resolution.
- Liquid reagents do not require low-temperature storage and can be stored at room temperature for more than 30 days.
- Power-off protection design, no data loss, and automatic recovery of working status when power is restored.
- Intelligent fault alarm and automatic reset to avoid abnormal rotation of the instrument.
- Micro reflux digestion technology and innovative automatic direct wind speed cooling device improve digestion and cooling efficiency and shorten sample processing time.
- Automatic verification and judgment function to meet the requirements of new standards.
- Customizable development of special functions and services to ensure ease of use and user-friendliness.
Technical parameters
Measurement range | 0-200/500/1000/2000/5000/10000/20000mg/L, the range can be customized according to customer requirements. |
Accuracy | ±5% |
Repeatability | ≤5% |
Quantitative lower limit | 15mg/L |
Resolution | 0.1mg/L |
24h low concentration drift | ±5mg/L |
24h high concentration drift | ±5%F.S |
Measurement mode | point measurement/cycle measurement/external signal trigger measurement |
Standard solution verification | immediate verification/automatic verification |
Instrument calibration | automatic calibration/manual calibration, calibration cycle and time can be set |
Automatic cleaning | cleaning cycle can be set |
Digestion time | can be chosen according to actual situation |
Data storage | can store data for 3 years |
Output | 2 channels of 4-20 mA analog output; RS232/RS485 digital output |
Power requirements | AC220±10%V,50±10%Hz,10A; power consumption less than 100W |
Operating temperature | (5~40)℃ |
Dimensions | 700mm(H) x 500mm(W) x 400(D)mm excluding the pretreatment system cabinet |
Weight | less than 30kg, excluding the pretreatment system cabinet. |