Three advantages
Double cleaning 30 days free maintenance
- Turbidity: automatic sewage discharge, automatic flushing
- Residual chlorine: double cleaning and enhanced cleaning effect (electrochemical cleaning + mechanical cleaning)
-Mechanical cleaning: the rotating impeller is driven by water to polish the platinum working electrode continuously to ensure the smooth surface of the working electrode.
-Electrochemical cleaning: The software controls WE-CE to conduct periodic electrochemical cleaning, by generating bubbles on the surface of the electrode and oxidizing organic insoluble materials on the surface of the electrode.
Three innovative designs to ensure the accuracy of the measurement data
- Aiming unstable flow rate will cause a large gap in the measurement data
Dongrun countermeasure: adopt the steady flow design to ensure the stable flow rate
- Aiming bubbles in water affects photorefraction data
Dongrun countermeasure: using steady current defoamizer to remove interference
- Specific to the environmental changes can affect the measurement data
Dongrun countermeasure: the residual chlorine of constant voltage three electrodes with pH, temperature automatic compensation
More economical, more environmentally friendly
- Cost saving, without expensive reagents / internal fluids
Constant voltage three-electrode residual chlorine, without film coating, expensive reagent / electrolyte
- Water saving, only 300ml/min
Series waterways are designed to replace multiple single waterways, with only 300ml of water per minute
- Convenient installation, small comprehensive space. 1 box body integrated multi-channel sensor, the comprehensive cost performance is better
Temperature | |
Measurement method | platinum resistance temperature sensor method |
Measuring range | -10-100℃ |
Resolution | 0.1℃ |
accuracy | ±0.2℃ |
Response time | <30 seconds |
MTBF | 1440 hours |
PH/ORP | |
Measurement method | glass electrode method / platinum electrode method |
Measuring range | 0.00~14.00pH -2000 ~ +2000mV |
Resolution | 0.01 pH/0.1mV |
Repeatability | 0.03pH |
accuracy | 0.1pH/5mV |
Response time | <30 seconds |
MTBF | 1440 hours |
Residual chlorine | |
Measurement method | constant voltage method or polarographic analysis method |
Measuring range | 0.02 ~ 2 mg/L constant voltage |
Resolution | 0.01 mg / L |
Repeatability | 0.5% constant voltage |
Response time | <60 seconds |
MTBF | 1440 hours |
Other indicators | |
Power supply | AC 220V± 10%, 50Hz |
Working environment | temperature: (0 ~ 50) ℃ |
Humidity | 95% RH (no condensation) |
Communication interface | RS485 Modbus RTU protocol |
Output interface | USB, available for data transfer storage |
Installation method | wall-mounted installation |
Water sample inlet | outer diameter 1 / 4, inch pipe |
Water sample drain | 3 / 8 inch pipe |
Dimensions | 480 * 360 * 230 (HWD) |